Download SPECT/CT Basics, Technology Updates, Quality Assurance, and Applications S. Cheenu Kappadath, PhD

January 15, 2018 | Author: Anonymous | Category: , Science, Health Science, Radiology
Share Embed


Short Description

Download Download SPECT/CT Basics, Technology Updates, Quality Assurance, and Applications S. Cheenu Kappadath, PhD...

Description

SPECT/CT

Basics, Technology Updates, Quality Assurance, and Applications S. Cheenu Kappadath, PhD Department of Imaging Physics University of Texas M D Anderson Cancer Center, Houston, Texas

[email protected]

Educational Objectives 1.

2.

3.

Understand the underlying principles of SPECT/CT image acquisition, processing and reconstruction Understand current and future clinical applications of SPECT/CT imaging Familiarization with commercially-available SPECT/CT systems

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Basics 

Single Photon Emission Computed Tomography 

 





Radio-pharmaceutical administration – injected, ingested, or inhaled Bio-distribution of pharmaceutical – uptake time Decay of radionuclide from within the patient – the source of information Gamma camera detects gamma rays and images (tomography) the radio-pharmaceutical distribution within the patient – SPECT Used for visualization of functional information based on the specific radio-pharmaceutical uptake mechanism

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Hardware Anatomy of a Gamma camera 1. 2. 3. 4. 5.

Collimator Scintillation Detector Photomultiplier Tubes Position Circuitry Data Analysis Computer

© U of British Columbia

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Back-Projection Model

g(s,) = f(x,y) along an in-plane line integral

© Bruyant, P. P., J Nucl Med 2002; 43:1343-1358

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Crystal Thickness 

Thinner crystals   spatial resolution    



interactions occur at a better defined depth multiple interactions less likely less light spread  interaction likelihood for higher energy ’s

Thicker crystals   sensitivity   

 interaction likelihood (esp. for higher E ’s)  likelihood of multiple interactions greater light spread   spatial resolution

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Spatial Resolution 

Intrinsic Spatial and Energy Resolution   

# of scintillation photons, N  Gamma-ray energy, E Spatial Resolution = 100  /N  1/N  1/E Energy Resolution = 100  FWHM/E  1/E

B Le H





Collimator Resolution System Resolution

S. Cheenu Kappadath, PhD

D ( Le  H  B ) Rg  Le

Rs2  Ri2  Rg2

Le D

AAPM 2009 - Anaheim, CA

SPECT Acquisitions 

SPECT acquires 2-D projections of a 3-D volume

© SPECT in the year 2000: Basic principles, JNMT 24:233, 2000 © Yale School of Medicine

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Radon transform angular symmetry violated in SPECT P() Anterior View

S. Cheenu Kappadath, PhD



P(+) horizontally flipped Posterior

AAPM 2009 - Anaheim, CA

Radon transform angular symmetry violated in SPECT  

Why ? Due to Differential Attenuation

L

b a i c I0 I(i+)

b

I(i)

I(i) = I0 e-a (L)dL c

I(i+) = I0 e-a (L)dL 

Other mediating factors:  

S. Cheenu Kappadath, PhD

distance-dependent resolution depth-dependent scatter AAPM 2009 - Anaheim, CA

SPECT Acquisitions  

SPECT projections acquired over 360° Exception: Cardiac SPECT acquired over 180° 

0° 0°



180°

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT images have isotropic voxel size 2-D filter of projections  3-D post-reconstruction filter No volume smoothing transverse

sagittal

coronal

Butterworth: 0.6 Nyquist, 10th order S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Filtered BackProjection 

FBP based on ideal Radon inversion formula 



SPECT imaging systems are neither angularly symmetric nor shift-invariant 



assumes a linear, shift-invariant system and angular symmetry of projections

SPECT projection data affected by attenuation, scatter, and spatial resolution that are all depth-or distance-dependent

Thus, FBP reconstruction cannot adequately model the physics of SPECT

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Conventional SPECT Corrections Attenuation: Chang post-processing algorithm L(x,y,i)

I(x,y) = SPECT image w/o AC I(x,y,i) = IAC(x,y).e-L(x,y,i) IAC (x,y) = I(x,y) / {(1/M).i

e-L(x,y,i)};

I(x,y,i) i

i = 1, M

IAC(x,y)

Scatter: Energy window subtraction

Lower Scatter Window

PhotoPeak Window

Upper Scatter Window STD in acrylic

20000

STD in air

Counts

P(x,y) = projections w/ scatter PLE(x,y) = projection at lower energy PHE(x,y) = projection at higher energy PSC (x,y) = P(x,y) – kL.PLE(x,y) – kH.PHE(x,y)

Energy Spectrum of Sm-153 30000

STD in acrylic with TEW Scatter Correction

10000

0 25

50

75

100

125

150

Photon Energy [keV]

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Iterative Reconstruction Maximum Likelihood-Expectation Maximization (ML-EM)  Accounts for the statistical nature of SPECT imaging  Incorporates the system response p(b,d) – the probability that a photon emitted from an object voxel b is detected by projection pixel

d

voxel b

detector d

 p(b,d) captures…

1. Depth-dependent resolution 2. Position-dependent scatter 3. Depth-dependent attenuation

 Use a measured attenuation map along with models of scatter and camera resolution to perform a far more accurate reconstruction S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Iterative Recon: Attenuation Modeling

b along a line integral … g(s,) = f(x,y) * pattn(x,y,s,) pattn(x,y,s,) = probability due to attenuation pattn(x,y,s,) = exp(-ab(x’,y’)x’,y’))

S. Cheenu Kappadath, PhD

a AAPM 2009 - Anaheim, CA

SPECT Iterative Recon: System Resolution Modeling Distance-dependent collimator beam ________ Rs =  Ri2 + Rc2 r Pencil Beam (FBP) Intrinsic Detector Resolution Ri

- iterative) Fan Beam (2D

Cone Beam (3D iterative) S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Iterative Recon: Resolution Modeling

2D: g(s,) = f(x,y) * pres(x,y,s,) 3D: g(s,) = f(x,y,z) * pres(x,y,z,s,) pres = probability due to resolution “fan of acceptance” (2D fan beam model) “cone of acceptance” (3D cone beam model)

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT Imaging: Scatter 

Scatter compensation occurs before attenuation  







the photopeak window contains scatter attenuation accounts for the removal of photopeak photons

Scatter contribution estimated as a weighted sum of one or more adjacent energy window images, Ci(x,y,) S(x,y,) = i ki × Ci(x,y,) Subtract scatter prior to reconstruction Pcorr(x,y,)  P(x,y,) - S(x,y,) Incorporate scatter into forward projection P(x,y,)  Pcorr(x,y,) + S(x,y,)

S. Cheenu Kappadath, PhD

SC techniques: DEW TEW ESSE

AAPM 2009 - Anaheim, CA

SPECT Iterative Reconstruction 



True projection intensity = sum of true voxel intensities weighted by detection probabilities

Forward Projection

True voxel intensity = sum of true detector intensities weighted by detection probabilities

Back Projection

S. Cheenu Kappadath, PhD

B

y (d )    (b) p (b, d ) b 1

D

 (b)   y (d ) p (b, d ) d 1

AAPM 2009 - Anaheim, CA

Iterative Reconstruction Flow Diagram D

 (b )  [k ]

 [ k 1] (b ) 

d 1

y ( d ) p (b , d ) [k ]   b '1 (b ') p (b ', d ) B

D

 p (b , d ) d 1

In clinical practice, the stopping criteria is number of iterations (a time constraint) instead of a convergence criteria.

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Ordered Subset EM (OSEM) 







Each OSEM iteration is a ML-EM iteration using an ordered subset of n (out of N) projections (eg: 4/36 views - 9 subsets, start with 0°,90°,180°,270° views) The next OSEM iteration starts with the result of the previous OSEM iterations but uses a different ordered subset of n projections (next set uses 10°,100°,190°,280° views)  rate of convergence by using an ordered subset of all N projections for each iteration m OSEM iterations with n subsets each  mn ML-EM iterations using all N each time

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

OSEM Iterative SPECT Reconstruction: Attenuation and Scatter Correction Un-Corrected

Corrected

Note the “hot-rim” artifact S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

OSEM Iterative SPECT Reconstruction: Collimator Resolution Modeling 99mTc

Bone Scan (osteosarcoma), LEHR Collimator

Standard Filtered Backprojection

2-D OSEM w/ fan beam modeling (m=12,n=10)

2-D pre-filter: Butterworth, fc = 0.6 Nyquist, order = 10

3-D OSEM w/ cone beam modeling (m=25,n=10) S. Cheenu Kappadath, PhD

3-D Gaussian Post-Filter (7.8 mm FWHM)

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT/CT Hybrid Imaging: Why? 

Non-uniform attenuation maps required 





Previous methods used constant  maps that work for brain but are problematic for thorax and pelvis radioactive source-based transmission CT – time penalty

Functional-anatomical overlay (image fusion)  

Improve localization of uptake regions Increase confidence in interpretation

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

CT-based AC for SPECT/CT CT

CTAC

μ‐map

CT noise  reduced

Smooth, re‐bin CT to match SPECT  Register CT w/ SPECT

Apply bi‐linear transform  on pixel‐by‐pixel basis

Reconstructed  SPECT Transition Matrix

aijk

S. Cheenu Kappadath, PhD

Other factors:  ‐SPECT projections ‐Scatter estimates ‐Collimator response AAPM 2009 - Anaheim, CA

CT-based  values Material attenuation versus Energy Air

Muscle

Bone

Photoelectric effect Compton scatter dominant dominant 

0.3





(cm2/g) 0.2



0.1 CT



0 0

100

200 Energy

S. Cheenu Kappadath, PhD

300

400

m = k ¥ CT-HU (simple but not accurate) Compton Scatter probability proportional to e- density Photoelectric effect probability proportional to (Z/E)3 Attenuation mismatch between PE and CS with energy for high Z

500

(keV) AAPM 2009 - Anaheim, CA

CT-based  values - HU-to-cm-1 conversion - not linearly related - piece-wise linear - bi- or tri-modal - Effective energy differences - CT (~ 70 – 80 keV) - SPECT (nuclide dependent) eg: 140 keV for Tc-99m

CT Number-to-Tc-99m  v alue Function 0.3

 value (cm-1)

0.25 0.2 0.15 0.1 0.05

1000

200

0

-1000

0

CT Number (HU)

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT/CT Hybrid Imaging: Iterative Reconstruction

FBP w/ Butterworth 0.4/5

99mTc

EC-DG (NSCLC)

3-D OSEM w/ resolution modeling

3-D OSEM w/ resolution and attenuation modeling S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT/CT QA/QC 

Planar (AAPM Reports 6 and 9; NEMA NU 1-1994)  



SPECT (AAPM Report 22 and 52)  



Inherently includes all planar gamma camera QA Energy/Spatial resolution, uniformity, deadtime, sensitivity, rotational uniformity, opposed-head registration, etc.

Uniformity and Contrast Resolution

SPECT/CT (AAPM TG 177: Jim Halama)  

NM-CT registration CT-HU to linear attenuation () transformation

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

NM-CT Registration 

Use Co-57 button sources w/ SPECT phantom

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

CT-HU to -map transformation 

Use an electron density phantom

CIRS Inc.

CT image: -790 to 235 HU

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Commercial SPECT/CT systems Siemens SymbiaT (1-, 2-, 6, 16-slice CT) GE Hawkeye (1- or 4-slice CT)

S. Cheenu Kappadath, PhD

Philips BrightView (Flat-panel CT)

AAPM 2009 - Anaheim, CA

GE – Millennium VG Hawkeye 

NM    



3/8” and 1” NaI(Tl) crystals 16 simultaneous energy windows Slip-ring gantry Body-contouring based on infrared-based transmitters

CT     

Co-planar, dental tube, 4-slice 20 mm beam no additional real estate needed Resolution: 3.5 or 1.75 mm (transaxial); 5 or 10 mm (axial) Time-averaged: 23 s per rotation (slow-scan) kVp: 120 – 140; mA: 1 – 2.5

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Phillips – BrightView XCT 

NM    



3/8” and ¾” NaI(Tl) crystals Energy-independent flood calibration (up to 300 keV) 15 simultaneous energy windows Body-contouring based on tissue impedance

CT     

Co-planar, flat-panel detector, 14 cm axial FOV no additional real estate needed High-resolution: 0.33 mm isotropic voxels Time-averaged: 12 s or 24 s per rotation (slow-scan) kVp: 120; mA: 5 – 80

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Siemens - SymbiaT 

NM    



3/8” and 5/8” NaI(Tl) crystals Energy-independent flood calibration (up to 300 keV) 6 simultaneous energy windows Body-contouring based on infrared-based transmitters

CT    

Diagnostic CT scanner kVp: 80/110/130; mA: 20 – 345 (T16) & 30 – 240 (T6) Scan time: 0.5, 0.6, 1, 1,5 s per rotation 1-, 2-, 6-, and 16-slice CT scanners

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Outline      

Review of SPECT principles Iterative SPECT reconstruction Hybrid SPECT/CT imaging SPECT/CT quality assurance Commercial SPECT/CT systems SPECT/CT clinical applications

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Clinical SPECT/CT Imaging 

Stress/Rest Myocardial Perfusion Imaging  

      

Stress: 99mTc-sestaMIBI or 99mTc-Tetrafosmin Rest: 99mTc-labeled agents or 201Tl-chloride

99mTc-MDP:

bone diseases, bone metasteses 99mTc-sestaMIBI: parathyroid adenomas 99mTc-sulphur colloid: liver/spleen, lymphoscintigraphy 111In-Pentetreotide: neuroendocrine cancers 111In-ProstaScint: prostate cancer 123I/131I-MIBG: pheochromocytoma, neuroblastoma 131I-NaI: thyroid cancer

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Clinical SPECT/CT Imaging      

99mTc-CEA:

colorectal cancer 99mTc-RBCs: hemangioma 99mTc-HMPAO, -ECD: brain perfusion 111In-WBC: infection 67Ga-citrate: inflammation, lymphoma 201Tl-chloride: tumor perfusion

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Clinical Benefits of SPECT/CT 

Visualization, diagnosis and interpretation of primary and metastatic diseases  

 



higher sensitivity and contrast than Planar imaging CT scan increases confidence in interpretation of SPECT examination

Surgical planning and IMRT treatment planning 90Y-microspheres radio-embolotherapy (selective internal RT or micro-brachytherapy) Internal radio-pharmaceutical therapy planning

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT/CT: Limitations 

Patient motion  



Contrast CT  



between SPECT and CT scans respiratory and cardiac motion during SPECT acquisitions contrast introduces electron density-material mismatch  map algorithms do not yet account for contrast CT

Absolute quantification (Bq/ml) not yet fully developed   

radionuclide-dependent acquisition/reconstruction technique-dependent calibration techniques not yet standardized

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

SPECT/CT: Future Applications     

Whole body SPECT/CT (analogous to PET/CT) Quantification of absolute activity (like PET) Compensation for CT contrast in  map Compensation for respiratory, cardiac motion SPECT/CT-based 3-D dosimetry/treatment planning

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Future: Whole-body Bone SPECT/CT 

Tc-99m MDP Bone Imaging

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA



99m-Tc MDP SPECT/CT: Fused Coronal views

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA

Future: Multi-nuclide SPECT/CT

 

Maximum Intensity Projection (MIP) of a dual-isotope (Tc-99m and I-123) SPECT/CT mouse study. Published by the Molecular Imaging Center for Excellence newsletter, SNM publication Volume 2, 2008

S. Cheenu Kappadath, PhD

AAPM 2009 - Anaheim, CA